SKU(재고 관리 코드):LMC031Hu
Magnetic Luminex Assay Kit for Interleukin 4 Receptor (IL4R) ,etc.
Magnetic Luminex Assay Kit for Interleukin 4 Receptor (IL4R) ,etc.
Product No.
LMC031Hu
Organism Species
Homo sapiens (Human).
Sample Type
Serum, plasma, tissue homogenates and other biological fluids
Test Method
Double-antibody Sandwich
Assay Length
3.5h
Detection Range
0.98-1000pg/mL
Sensitivity
The minimum detectable dose of this kit is typically less than 0.327 pg/mL.
UOM
1Plex 2Plex 3Plex 4Plex 5Plex 6Plex 7Plex 8Plex
Specificity
This assay has high sensitivity and excellent specificity for detection of Interleukin 4 Receptor (IL4R) ,etc..
No significant cross-reactivity or interference between Interleukin 4 Receptor (IL4R) ,etc. and analogues was observed.
Recovery
Matrices listed below were spiked with certain level of recombinant Interleukin 4 Receptor (IL4R) ,etc. and the recovery rates were calculated by comparing the measured value to the expected amount of Interleukin 4 Receptor (IL4R) ,etc. in samples.
Matrix | Recovery range (%) | Average(%) |
serum(n=5) | 94-101 | 97 |
EDTA plasma(n=5) | 92-101 | 95 |
heparin plasma(n=5) | 83-99 | 90 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Interleukin 4 Receptor (IL4R) ,etc. were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Interleukin 4 Receptor (IL4R) ,etc. were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Interleukin 4 Receptor (IL4R) ,etc. and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample | 1:2 | 1:4 | 1:8 | 1:16 |
serum(n=5) | 80-101% | 81-94% | 81-95% | 92-101% |
EDTA plasma(n=5) | 79-92% | 97-105% | 96-105% | 92-105% |
heparin plasma(n=5) | 78-93% | 90-97% | 82-91% | 85-103% |
Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Reagents and materials provided
Reagents | Quantity | Reagents | Quantity |
96-well plate | 1 | Plate sealer for 96 wells | 4 |
Pre-Mixed Standard | 2 | Standard Diluent | 1×20mL |
Pre-Mixed Magnetic beads (22#:IL4R) | 1 | Analysis buffer | 1×20mL |
Pre-Mixed Detection Reagent A | 1×120μL | Assay Diluent A | 1×12mL |
Detection Reagent B (PE-SA) | 1×120μL | Assay Diluent B | 1×12mL |
Sheath Fluid | 1×10mL | Wash Buffer (30 × concentrate) | 1×20mL |
Instruction manual | 1 |
Assay procedure summary
1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.
Magazine | Citations |
Int J Nanomedicine | Treating glioblastoma multiforme with selective high-dose liposomal doxorubicin chemotherapy induced by repeated focused ultrasound PubMed: PMC3289450 |
Multidisciplinary Respiratory Medicine | Role of interleukin 4 and its receptor in clinical presentation of chronic extrinsic allergic alveolitis: a pilot study PubMed: PMC3682858 |
The Open Respiratory Medicine Journal | Biomarkers of Fibroproliferative Healing in Fibrosing Idiopathic Interstitial Pneumonias PubMed: PMC3551240 |
Exp Ther Med. | PAR2, IL4R, TGFβ and TNFα in bronchoalveolar lavage distinguishes extrinsic allergic alveolitis from sarcoidosis Pubmed:Pmc4079423 |
diagnostics | Interstitial Score and Concentrations of IL-4R¦Á, PAR-2, and MMP-7 in Bronchoalveolar Lavage Fluid Could Be Useful Markers for Distinguishing Idiopathic Interstitial?¡ 33924683 |
Catalog No. | Related products for research use of Homo sapiens (Human) Organism species | Applications (RESEARCH USE ONLY!) |
RPC031Hu01 | Recombinant Interleukin 4 Receptor (IL4R) | Positive Control; Immunogen; SDS-PAGE; WB. |
PAC031Hu01 | Polyclonal Antibody to Interleukin 4 Receptor (IL4R) | WB; IHC; ICC; IP. |
RAC031Hu21 | Recombinant Antibody to Interleukin 4 Receptor (IL4R) | WB; IF; ICC; IHC; IP; FCM. |
MAC031Hu22 | Monoclonal Antibody to Interleukin 4 Receptor (IL4R) | WB; IHC; ICC; IP. |
SEC031Hu | ELISA Kit for Interleukin 4 Receptor (IL4R) | Enzyme-linked immunosorbent assay for Antigen Detection. |
LMC031Hu | Magnetic Luminex Assay Kit for Interleukin 4 Receptor (IL4R) ,etc. | Magnetic Luminex Assay for Antigen Detection. |