제품 정보로 건너뛰기
1 1

Cloud-clone

SKU(재고 관리 코드):LMA170Ra

Magnetic Luminex Assay Kit for Activin B (ACVB) ,etc.

Magnetic Luminex Assay Kit for Activin B (ACVB) ,etc.

Detection range
Size
Sensitivity

Product No.

LMA170Ra

Organism Species

Rattus norvegicus (Rat).

Sample Type

Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids

Test Method

Competitive Inhibition

Assay Length

1.5h

Detection Range

0.98-1000pg/mL

Sensitivity

The minimum detectable dose of this kit is typically less than 0.327 pg/mL.

UOM

1Plex 2Plex 3Plex 4Plex 5Plex 6Plex 7Plex 8Plex

 

Specificity

This assay has high sensitivity and excellent specificity for detection of Activin B (ACVB) ,etc..
No significant cross-reactivity or interference between Activin B (ACVB) ,etc. and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Activin B (ACVB) ,etc. and the recovery rates were calculated by comparing the measured value to the expected amount of Activin B (ACVB) ,etc. in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 93-101 97
EDTA plasma(n=5) 99-105 102
heparin plasma(n=5) 82-95 92

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Activin B (ACVB) ,etc. were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Activin B (ACVB) ,etc. were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Activin B (ACVB) ,etc. and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 94-102% 82-94% 93-102% 80-96%
EDTA plasma(n=5) 90-97% 87-95% 83-104% 78-98%
heparin plasma(n=5) 92-103% 94-105% 78-97% 95-102%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:ACVB) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 50μL standard or sample to each well,
    add 10μL magnetic beads,and 50μL Detection Reagent A,incubate 60min at 37°C on shaker;
3. Wash plate on magnetic frame for three times;
4. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
5. Wash plate on magnetic frame for three times;
6. Add 100μL sheath solution, swirl for 2 minutes, read on the machine. 

     

    Magazine Citations
    Gastroenterology Research and Practice Serum Activins and Follistatin during the Treatment of Chronic Hepatitis C Genotypes 1 and 4 and Their Correlations with Viral Load and Liver Enzymes: A Preliminary Report. Pubmed: 24799891
    Gastroenterol Res Pract.? Serum activins and follistatin during the treatment of chronic hepatitis c genotypes 1 and 4 and their correlations with viral load and liver enzymes: A preliminary report Pubmed:Pmc3995172
    Clin Exp Med.? Effects?of?chronic hepatitis C?genotype?1?and 4 on serum activins and follistatin in treatment na?ve patients and their correlations with interleukin-6, tumour necrosis factor-α, viral load and liver damage. Pubmed:24925642
    Clin Biochem Serum activin B concentration as predictive biomarker for ectopic pregnancy Pubmed:26968108
    BMC Cancer. Activins and their related proteins in colon carcinogenesis: insights from early and advanced azoxymethane rat models of colon cancer. pubmed:27835986
    Australasian Journal of Ultrasound in Medicine The potential value of activin B and fibronectin for the triage of pregnancies of unknown location and prediction of first trimester viability 10.1002:ajum.12090
    Pathology & Oncology Research Profiling Activins and Follistatin in Colorectal Cancer According to Clinical Stage, Tumour Sidedness and Smad4 Status 34867090
    Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
    CEA170Ra ELISA Kit for Activin B (ACVB) Enzyme-linked immunosorbent assay for Antigen Detection.
    LMA170Ra Magnetic Luminex Assay Kit for Activin B (ACVB) ,etc. Magnetic Luminex Assay for Antigen Detection.
    전체 세부 정보 보기