FineTest
SKU(재고 관리 코드):ECA0003
Canine BNP(Brain Natriuretic Peptide) ELISA Kit
Canine BNP(Brain Natriuretic Peptide) ELISA Kit
The Canine BNP (Brain Natriuretic Peptide) ELISA Kit is a specialized research tool designed for the quantitative measurement of Brain Natriuretic Peptide in canine samples. BNP is a critical biomarker for heart-related conditions, and this kit enables researchers and clinicians to accurately assess BNP levels in canine biological samples. Key Features: High Specificity: This kit is specifically designed to detect canine BNP, ensuring minimal cross-reactivity with other molecules and reliable results. Quantitative Measurements: Provides precise quantitative data for comparisons and data analysis. Versatile Application: Suitable for use in various canine sample types, such as serum and plasma, accommodating a wide range of research and clinical applications. User-Friendly: The kit includes all necessary components and detailed instructions for easy and efficient experimentation. Reliability: Rigorous quality control processes and calibration standards ensure the consistency and accuracy of results. Empower your research on canine heart health and related conditions with the Canine BNP (Brain Natriuretic Peptide) ELISA Kit.
Product Name
Canine BNP(Brain Natriuretic Peptide) ELISA Kit
Alias
BNP ELISA Kit, Brain Natriuretic Peptide ELISA Kit, NPPB ELISA Kit, GC-B ELISA Kit, B-Type Natriuretic Peptide ELISA Kit, Natriuretic Peptide Precursor B ELISA Kit
Detection Method
Sandwich ELISA, Double Antibody
Application
BNP ELISA Kit allows for the in vitro quantitative determination of BNP concentrations in serum, plasma, tissue homogenates and other biological fluids.
Size
96T
Range
31.25-2000pg/ml
Sensitivity
< 18.75pg/ml
Species
Canine
UniProt ID
Standard Curve

Storage
2-8 ℃ for 6 months
Recovery
Matrices listed below were spiked with certain level of BNP and the recovery rates were calculated by comparing the measured value to the expected amount of BNP in samples.
|
Matrix |
Recovery range(%) |
Average(%) |
|
serum(n=5) |
87-102 |
98 |
|
EDTA plasma(n=5) |
86-99 |
95 |
|
heparin plasma(n=5) |
87-105 |
94 |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of BNP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
|
Sample |
1:2 |
1:4 |
1:8 |
|
serum(n=5) |
85-101% |
85-102% |
85-99% |
|
EDTA plasma(n=5) |
83-101% |
86-100% |
83-95% |
|
heparin plasma(n=5) |
81-97% |
85-100% |
86-100% |
CV(%)
Intra-Assay: CV<8%
Inter-Assay: CV<10%
Note
For Research Use Only
Share