Overview
| Product name: | Dog hs-cTnI(High Sensitivity Cardiac Troponin I) ELISA Kit |
| Reactivity: | Dog |
| Alternative Names: | High Sensitivity Cardiac Troponin I |
| Assay Type: | Sandwich |
| Sensitivity: | 4.94 pg/mL |
| Standard: | 1000 pg/mL |
| Range: | 15.63-1000 pg/mL |
| Sample Type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Cardiovascular biology |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Dog hs-cTnI. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Dog hs-cTnI. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Dog hs-cTnI, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Dog hs-cTnI in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.056 | 1.973 |
| 500.00 | 1.613 | 1.530 |
| 250.00 | 1.088 | 1.005 |
| 125.00 | 0.841 | 0.758 |
| 62.50 | 0.457 | 0.374 |
| 31.25 | 0.357 | 0.274 |
| 15.63 | 0.241 | 0.158 |
| 0.00 | 0.083 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant hs-cTnI and the recovery rates were calculated by comparing the measured value to the expected amount of hs-cTnI in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-92% | 86% |
| EDTA plasma(n=5) | 92-106% | 99% |
| Heparin plasma(n=5) | 87-99% | 93% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of hs-cTnI and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-104% | 82-96% | 92-105% | 83-92% |
| EDTA plasma(n=5) | 95-103% | 99-106% | 78-92% | 83-97% |
| Heparin plasma(n=5) | 87-95% | 78-91% | 97-103% | 95-101% |