SKU(재고 관리 코드):ELK9575

General AcCoA(Acetyl CoA) ELISA Kit

General AcCoA(Acetyl CoA) ELISA Kit

Sensitivity

General AcCoA(Acetyl CoA) ELISA Kit

Overview

Product name: AcCoA(Acetyl CoA) ELISA Kit
Reactivity: General
Alternative Names: AcCoA
Assay Type: Sandwich
Sensitivity: 0.119 nmol/mL
Standard: 20 nmol/mL
Detection Range: 0.32-20 nmol/mL
Sample Type: Cell Lysate, Tissue Lysate
Assay Length: 3.5h
Research Area: Metabolic pathway; Signal Transduction; Cancer
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to AcCoA. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to AcCoA. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain AcCoA, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of AcCoA in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
20.00 2.096 2.028
10.00 1.429 1.361
5.00 1.185 1.117
2.50 0.848 0.780
1.25 0.453 0.385
0.63 0.323 0.255
0.32 0.213 0.145
0.00 0.068 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant AcCoA and the recovery rates were calculated by comparing the measured value to the expected amount of AcCoA in samples.
Matrix Recovery range Average
serum(n=5) 86-99% 92%
EDTA plasma(n=5) 87-99% 93%
Heparin plasma(n=5) 80-95% 87%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of AcCoA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 81-92% 87-101% 89-99% 85-95%
EDTA plasma(n=5) 93-102% 93-106% 87-102% 84-98%
Heparin plasma(n=5) 89-102% 82-96% 88-97% 85-92%