SKU(재고 관리 코드):ELK9646
General F-E3(Free Estriol) ELISA Kit
General F-E3(Free Estriol) ELISA Kit
General F-E3(Free Estriol) ELISA Kit
Overview
Product name: | F-E3(Free Estriol) ELISA Kit |
Reactivity: | General |
Alternative Names: | F-E3; Free Estriol |
Assay Type: | Competitive Inhibition |
Sensitivity: | 0.218 ng/mL |
Standard: | 40 ng/mL |
Detection Range: | 0.63-40 ng/mL |
Sample Type: | Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
Assay Length: | 2h |
Research Area: | Endocrinology;Reproductive science;Hormone metabolism; |
Test principle: | This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with F-E3 protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to F-E3. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of F-E3 in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
40.00 | 0.215 | — |
20.00 | 0.310 | — |
10.00 | 0.596 | — |
5.00 | 0.895 | — |
2.50 | 1.169 | — |
1.25 | 1.585 | — |
0.63 | 1.869 | — |
0.00 | 2.157 | — |
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant F-E3 and the recovery rates were calculated by comparing the measured value to the expected amount of F-E3 in samples.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 83-95% | 89% |
EDTA plasma(n=5) | 79-93% | 86% |
Heparin plasma(n=5) | 92-105% | 98% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of F-E3 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-93% | 91-103% | 87-98% | 92-99% |
EDTA plasma(n=5) | 89-99% | 83-96% | 98-104% | 88-101% |
Heparin plasma(n=5) | 87-96% | 92-105% | 89-102% | 92-101% |