1
/
의
1
ELK Biotechnology
SKU(재고 관리 코드):ELK9584
General F-TESTO(FreeTestosterone) ELISA Kit
General F-TESTO(FreeTestosterone) ELISA Kit
General F-TESTO(FreeTestosterone) ELISA Kit
Overview
| Product name: | F-TESTO(FreeTestosterone) ELISA Kit |
| Reactivity: | General |
| Alternative Names: | F-TESTO |
| Assay Type: | Competitive Inhibition |
| Sensitivity: | 18.75 pg/mL |
| Standard: | 2000 pg/mL |
| Detection Range: | 31.25-2000 pg/mL |
| Sample Type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
| Assay Length: | 2h |
| Research Area: | Endocrinology;Reproductive science;Hormone metabolism; |
| Test principle: | This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with F-TESTO protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to F-TESTO. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of F-TESTO in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 0.198 | — |
| 1000.00 | 0.385 | — |
| 500.00 | 0.621 | — |
| 250.00 | 0.842 | — |
| 125.00 | 1.275 | — |
| 62.50 | 1.565 | — |
| 31.25 | 1.787 | — |
| 0.00 | 2.057 | — |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant F-TESTO and the recovery rates were calculated by comparing the measured value to the expected amount of F-TESTO in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-95% | 88% |
| EDTA plasma(n=5) | 83-95% | 89% |
| Heparin plasma(n=5) | 89-103% | 96% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of F-TESTO and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-104% | 87-103% | 91-102% | 87-102% |
| EDTA plasma(n=5) | 87-96% | 92-101% | 85-98% | 97-104% |
| Heparin plasma(n=5) | 92-101% | 82-96% | 95-102% | 95-104% |