제품 정보로 건너뛰기
1 1

ELK Biotechnology

SKU(재고 관리 코드):ELK9541

Goat vWF(Von Willebrand Factor) ELISA Kit

Goat vWF(Von Willebrand Factor) ELISA Kit

Sensitivity

Goat vWF(Von Willebrand Factor) ELISA Kit

Overview

Product name: Goat vWF(Von Willebrand Factor) ELISA Kit
Reactivity: Goat
Alternative Names: F8VWF; VWD; von Willebrand antigen 2
Assay Type: Competitive Inhibition
Sensitivity: 4.11 ng/mL
Standard: 800 ng/mL
Detection Range: 12.5-800 ng/mL
Sample Type: Plasma.
Assay Length: 2h
Research Area: Signal transduction;Hematology;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Goat vWF protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Goat vWF. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Goat vWF in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
800.00 0.191
400.00 0.387
200.00 0.593
100.00 0.967
50.00 1.156
25.00 1.585
12.50 1.877
0.00 2.132

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant vWF and the recovery rates were calculated by comparing the measured value to the expected amount of vWF in samples.
Matrix Recovery range Average
serum(n=5) 82-95% 88%
EDTA plasma(n=5) 90-105% 97%
Heparin plasma(n=5) 93-107% 100%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of vWF and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 97-106% 92-101% 82-90% 83-92%
EDTA plasma(n=5) 89-98% 86-98% 80-93% 81-98%
Heparin plasma(n=5) 89-102% 86-97% 95-103% 93-102%
전체 세부 정보 보기