Hamster CYP7A1(Cytochrome P450 7A1) ELISA Kit

Hamster CYP7A1(Cytochrome P450 7A1) ELISA Kit

Sensitivity

Hamster CYP7A1(Cytochrome P450 7A1) ELISA Kit

Overview

Product name: Hamster CYP7A1(Cytochrome P450 7A1) ELISA Kit
Reactivity: Hamster
Alternative Names: CYP7; CYP7-A1; CYPVII; Cholesterol 7 Alpha Hydroxylase; Cytochrome P450 Family 7 Subfamily A Polypeptide 1
Assay Type: Sandwich
Sensitivity: 0.133 ng/mL
Standard: 20 ng/mL
Detection Range: 0.32-20 ng/mL
Sample Type: Serum, plasma, tissue homogenates and other biological fluids
Assay Length: 3.5h
Research Area: Metabolic pathway;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Hamster CYP7A1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Hamster CYP7A1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Hamster CYP7A1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Hamster CYP7A1 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
20.00 2.135 2.050
10.00 1.638 1.553
5.00 1.188 1.103
2.50 0.933 0.848
1.25 0.532 0.447
0.63 0.387 0.302
0.32 0.179 0.094
0.00 0.085 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant CYP7A1 and the recovery rates were calculated by comparing the measured value to the expected amount of CYP7A1 in samples.
Matrix Recovery range Average
serum(n=5) 80-94% 87%
EDTA plasma(n=5) 95-107% 101%
Heparin plasma(n=5) 82-98% 90%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CYP7A1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 88-96% 90-101% 93-102% 88-102%
EDTA plasma(n=5) 86-92% 81-93% 83-96% 91-103%
Heparin plasma(n=5) 96-105% 89-98% 81-94% 87-98%