1
/
의
1
ELK Biotechnology
SKU(재고 관리 코드):ELK8060
Human Anti-CRP(Anti-C Reactive Protein Antibody) ELISA Kit
Human Anti-CRP(Anti-C Reactive Protein Antibody) ELISA Kit
Human Anti-CRP(Anti-C Reactive Protein Antibody) ELISA Kit
Overview
| Product name: | Human Anti-CRP(Anti-C Reactive Protein Antibody) ELISA Kit |
| Reactivity: | Human |
| Alternative Names: | C-RP; PTX1; Pentraxin-Related; Pentraxin 1 |
| Assay Type: | Competitive Inhibition |
| Sensitivity: | 1.14 ng/mL |
| Standard: | 200 ng/mL |
| Detection Range: | 3.13-200 ng/mL |
| Sample Type: | Serum, plasma and other biological fluids |
| Assay Length: | 2.5h |
| Research Area: | Tumor immunity;Infection immunity;Cardiovascular biology;Rheumatology; |
| Test principle: | This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Human Anti-CRP protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human Anti-CRP. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human Anti-CRP in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 0.212 | — |
| 100.00 | 0.386 | — |
| 50.00 | 0.583 | — |
| 25.00 | 0.833 | — |
| 12.50 | 1.156 | — |
| 6.25 | 1.511 | — |
| 3.13 | 1.697 | — |
| 0.00 | 2.226 | — |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant Anti-CRP and the recovery rates were calculated by comparing the measured value to the expected amount of Anti-CRP in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-97% | 88% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 83-95% | 89% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Anti-CRP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-98% | 92-104% | 82-96% | 95-102% |
| EDTA plasma(n=5) | 92-104% | 82-96% | 95-102% | 87-98% |
| Heparin plasma(n=5) | 86-97% | 90-98% | 80-93% | 92-105% |
Share