Human CAMP(Cathelicidin Antimicrobial Peptide) ELISA Kit
Human CAMP(Cathelicidin Antimicrobial Peptide) ELISA Kit
Human CAMP(Cathelicidin Antimicrobial Peptide) ELISA Kit
Overview
| Product name: | Human LL-37(Antibacterial Protein LL-37) ELISA Kit |
| Reactivity: | Human |
| Alternative Names: | CAP18; FALL39; HSD26; LL37; 18 kDa Cationic Antimicrobial Protein; Antibacterial protein FALL-39; FALL-39 peptide antibiotic; CAMP; Cathelicidin Antimicrobial Peptide |
| Assay Type: | Sandwich |
| Sensitivity: | 0.27 ng/mL |
| Standard: | 50 ng/mL |
| Range: | 0.79-50 ng/mL |
| Sample Type: | serum, plasma and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Infection immunity; |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human LL-37. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human LL-37. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human LL-37, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human LL-37 in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.054 | 1.971 |
| 25.00 | 1.587 | 1.504 |
| 12.50 | 1.027 | 0.944 |
| 6.25 | 0.866 | 0.783 |
| 3.13 | 0.509 | 0.426 |
| 1.57 | 0.335 | 0.252 |
| 0.79 | 0.228 | 0.145 |
| 0.00 | 0.083 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant LL-37 and the recovery rates were calculated by comparing the measured value to the expected amount of LL-37 in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-94% | 88% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 82-94% | 88% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of LL-37 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-98% | 83-94% | 85-98% | 85-97% |
| EDTA plasma(n=5) | 98-105% | 92-101% | 85-105% | 87-98% |
| Heparin plasma(n=5) | 85-92% | 79-96% | 89-100% | 85-96% |