SKU(재고 관리 코드):ELK9587
Human N-MID-OT(N-MIDOsteocalcin) ELISA Kit
Human N-MID-OT(N-MIDOsteocalcin) ELISA Kit
Human N-MID-OT(N-MIDOsteocalcin) ELISA Kit
Overview
Product name: | Human N-MID-OT(N-MIDOsteocalcin) ELISA Kit |
Reactivity: | Human |
Alternative Names: | N-MID-OT |
Assay Type: | Sandwich |
Sensitivity: | 0.38 ng/mL |
Standard: | 40 ng/mL |
Detection Range: | 0.63-40 ng/mL |
Sample Type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
Assay Length: | 3.5h |
Research Area: | Metabolic pathway;Endocrinology;Hormone metabolism;Bone metabolism; |
Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human N-MID-OT. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human N-MID-OT. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human N-MID-OT, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human N-MID-OT in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
40.00 | 2.258 | 2.156 |
20.00 | 1.625 | 1.523 |
10.00 | 1.183 | 1.081 |
5.00 | 0.826 | 0.724 |
2.50 | 0.518 | 0.416 |
1.25 | 0.357 | 0.255 |
0.63 | 0.236 | 0.134 |
0.00 | 0.102 | 0.000 |
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant N-MID-OT and the recovery rates were calculated by comparing the measured value to the expected amount of N-MID-OT in samples.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 95-107% | 101% |
EDTA plasma(n=5) | 89-103% | 96% |
Heparin plasma(n=5) | 83-99% | 95% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of N-MID-OT and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 82-97% | 88-105% | 79-93% | 87-98% |
EDTA plasma(n=5) | 89-96% | 79-98% | 79-91% | 90-99% |
Heparin plasma(n=5) | 85-96% | 87-98% | 95-104% | 93-102% |