1
/
의
1
ELK Biotechnology
SKU(재고 관리 코드):ELK3376
Human NKX6-1(NK6 Homeobox Protein 1) ELISA Kit
Human NKX6-1(NK6 Homeobox Protein 1) ELISA Kit
Human NKX6-1(NK6 Homeobox Protein 1) ELISA Kit
Overview
| Product name: | Human NKX6-1(NK6 Homeobox Protein 1) ELISA Kit |
| Reactivity: | Human |
| Alternative Names: | NKX6A; Nkx6.1; NK6 Transcription Factor Related,Locus 1; Homeobox protein NK-6 homolog A |
| Assay Type: | Sandwich |
| Sensitivity: | 0.131 ng/mL |
| Standard: | 20 ng/mL |
| Detection Range: | 0.32-20 ng/mL |
| Sample Type: | Tissue homogenates, cell lysates and other biological fluids. |
| Assay Length: | 3.5h |
| Research Area: | Signal transduction; |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human NKX6-1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human NKX6-1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human NKX6-1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human NKX6-1 in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.136 | 2.035 |
| 10.00 | 1.657 | 1.556 |
| 5.00 | 1.035 | 0.934 |
| 2.50 | 0.834 | 0.733 |
| 1.25 | 0.583 | 0.482 |
| 0.63 | 0.327 | 0.226 |
| 0.32 | 0.230 | 0.129 |
| 0.00 | 0.101 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant NKX6-1 and the recovery rates were calculated by comparing the measured value to the expected amount of NKX6-1 in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-91% | 84% |
| EDTA plasma(n=5) | 80-96% | 88% |
| Heparin plasma(n=5) | 88-101% | 94% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of NKX6-1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-98% | 92-105% | 95-107% | 82-96% |
| EDTA plasma(n=5) | 88-104% | 84-98% | 95-105% | 88-102% |
| Heparin plasma(n=5) | 78-92% | 83-96% | 95-107% | 87-102% |
Share