SKU(재고 관리 코드):ELK9590
Human TRACP-5b(Tartrate Resistant Acid Phosphatase 5b) ELISA Kit
Human TRACP-5b(Tartrate Resistant Acid Phosphatase 5b) ELISA Kit
Human TRACP-5b(Tartrate Resistant Acid Phosphatase 5b) ELISA Kit
Overview
Product name: | Human TRACP-5b(Tartrate Resistant Acid Phosphatase 5b) ELISA Kit |
Reactivity: | Human |
Alternative Names: | ACP5;TRAP5b;SPENCDI |
Assay Type: | Sandwich |
Sensitivity: | 0.47 ng/mL |
Standard: | 50 ng/mL |
Detection Range: | 0.78-50 ng/mL |
Sample Type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
Assay Length: | 3.5h |
Research Area: | Signal Transduction;Signal Transduction;Metabolism |
Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human TRACP-5b. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human TRACP-5b. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human TRACP-5b, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human TRACP-5b in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
50.00 | 2.204 | 2.117 |
25.00 | 1.543 | 1.456 |
12.50 | 1.173 | 1.086 |
6.25 | 0.907 | 0.820 |
3.13 | 0.525 | 0.438 |
1.57 | 0.369 | 0.282 |
0.79 | 0.267 | 0.180 |
0.00 | 0.087 | 0.000 |
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant TRACP-5b and the recovery rates were calculated by comparing the measured value to the expected amount of TRACP-5b in samples.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 90-105% | 97% |
EDTA plasma(n=5) | 96-107% | 102% |
Heparin plasma(n=5) | 80-93% | 86% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of TRACP-5b and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 95-102% | 95-103% | 88-96% | 81-95% |
EDTA plasma(n=5) | 79-96% | 95-102% | 82-97% | 88-105% |
Heparin plasma(n=5) | 79-93% | 87-98% | 95-103% | 84-98% |