Product name: | Mouse Ang1-7(Angiotensin1-7) ELISA Kit |
Reactivity: | Mouse |
Alternative Names: | Angiotensin 1-7 |
Assay Type: | Sandwich |
Sensitivity: | 18.75 pg/mL |
Standard: | 2000 pg/mL |
Detection Range: | 31.25-2000 pg/mL |
Sample Type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
Assay Length: | 3.5h |
Research Area: | Cardiovascular;Signal Transduction;Metabolism |
Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse Ang1-7. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse Ang1-7. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse Ang1-7, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse Ang1-7 in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
2000.00 | 1.972 | 1.875 |
1000.00 | 1.663 | 1.566 |
500.00 | 1.108 | 1.011 |
250.00 | 0.876 | 0.779 |
125.00 | 0.576 | 0.479 |
62.50 | 0.385 | 0.288 |
31.25 | 0.226 | 0.129 |
0.00 | 0.097 | 0.000 |
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant Ang1-7 and the recovery rates were calculated by comparing the measured value to the expected amount of Ang1-7 in samples.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-95% | 87% |
EDTA plasma(n=5) | 84-96% | 90% |
Heparin plasma(n=5) | 83-95% | 89% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Ang1-7 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-97% | 85-94% | 84-98% | 86-93% |
EDTA plasma(n=5) | 88-92% | 86-97% | 90-101% | 92-102% |
Heparin plasma(n=5) | 85-97% | 95-107% | 85-94% | 82-96% |