1
/
의
1
ELK Biotechnology
SKU(재고 관리 코드):ELK9754
Mouse Chia(Acidic mammalian chitinase) ELISA Kit
Mouse Chia(Acidic mammalian chitinase) ELISA Kit
Mouse Chia(Acidic mammalian chitinase) ELISA Kit
Overview
| Product name: | Mouse Chia(Acidic mammalian chitinase) ELISA Kit |
| Reactivity: | Mouse |
| Alternative Names: | Acidic mammalian chitinase [Precursor]; Acidic mammalian chitinase; AMCase; Chia; CHIA_HUMAN; CHIT2; chitinase, acidic; DKFZp313J1722; ECF L; Lung specific protein TSA1902; Lung-specific protein TSA1902; RP5 1125M8.1; TSA1902 |
| Assay Type: | Sandwich |
| Sensitivity: | 0.058 ng/mL |
| Standard: | 10 ng/mL |
| Detection Range: | 0.16-10 ng/mL |
| Sample Type: | serum, plasma, saliva, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Immunology; Microbiology; Signal Transduction |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse Chia. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse Chia. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse Chia, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse Chia in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.952 | 1.870 |
| 5.00 | 1.611 | 1.529 |
| 2.50 | 1.182 | 1.100 |
| 1.25 | 0.963 | 0.881 |
| 0.63 | 0.475 | 0.393 |
| 0.32 | 0.389 | 0.307 |
| 0.16 | 0.215 | 0.133 |
| 0.00 | 0.082 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant Chia and the recovery rates were calculated by comparing the measured value to the expected amount of Chia in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-96% | 87% |
| EDTA plasma(n=5) | 86-96% | 91% |
| Heparin plasma(n=5) | 84-96% | 90% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Chia and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-100% | 84-98% | 95-101% | 88-97% |
| EDTA plasma(n=5) | 79-92% | 81-97% | 87-95% | 85-94% |
| Heparin plasma(n=5) | 96-105% | 92-103% | 90-99% | 89-102% |
Share