SKU(재고 관리 코드):ELK9604
Mouse E-Cad(E-Cadherin) ELISA Kit
Mouse E-Cad(E-Cadherin) ELISA Kit
Mouse E-Cad(E-Cadherin) ELISA Kit
Overview
Product name: | Mouse E-Cad(E-Cadherin) ELISA Kit |
Reactivity: | Mouse |
Alternative Names: | CDH1; Arc-1; CD324; CDHE; LCAM; UVO; CAM 120/80; Epithelial Cadherin; Uvomorulin |
Assay Type: | Sandwich |
Sensitivity: | 37.50 pg/mL |
Standard: | 4000 pg/mL |
Detection Range: | 62.50-4000 pg/mL |
Sample Type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
Assay Length: | 3.5h |
Research Area: | Signal Transduction;Cancer;Developmental Biology |
Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse E-Cad. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse E-Cad. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse E-Cad, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse E-Cad in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
4000.00 | 2.179 | 2.091 |
2000.00 | 1.607 | 1.519 |
1000.00 | 1.140 | 1.052 |
500.00 | 0.852 | 0.764 |
250.00 | 0.527 | 0.439 |
125.00 | 0.305 | 0.217 |
62.50 | 0.252 | 0.164 |
0.00 | 0.088 | 0.000 |
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant E-Cad and the recovery rates were calculated by comparing the measured value to the expected amount of E-Cad in samples.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 84-96% | 90% |
EDTA plasma(n=5) | 96-107% | 101% |
Heparin plasma(n=5) | 83-95% | 89% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of E-Cad and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-98% | 93-102% | 89-97% | 85-94% |
EDTA plasma(n=5) | 95-102% | 87-103% | 89-102% | 87-102% |
Heparin plasma(n=5) | 85-94% | 87-101% | 87-101% | 82-93% |