제품 정보로 건너뛰기
1 1

ELK Biotechnology

SKU(재고 관리 코드):ELK9400

Mouse sFLT(Soluble Fms-Like Tyrosine Kinase Receptor 1) ELISA Kit

Mouse sFLT(Soluble Fms-Like Tyrosine Kinase Receptor 1) ELISA Kit

Sensitivity

Mouse sFLT(Soluble Fms-Like Tyrosine Kinase Receptor 1) ELISA Kit

Overview

Product name: Mouse sFLT(Soluble Fms-Like Tyrosine Kinase Receptor 1) ELISA Kit
Reactivity: Mouse
Alternative Names: Flt1; Fms-Related Tyrosine Kinase 1; Vascular Permeability Factor Receptor; Fms-like tyrosine kinase 1; Tyrosine-protein kinase receptor FLT
Assay Type: Sandwich
Sensitivity: 29 pg/mL
Standard: 5000 pg/mL
Detection Range: 78.13-5000 pg/mL
Sample Type: serum, plasma, tissue homogenates, saliva, cell lysates, follicular fluid and other biological fluids
Assay Length: 3.5h
Research Area: Signal transduction;Tumor immunity;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse sFLT. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse sFLT. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse sFLT, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse sFLT in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
5000.00 2.201 2.103
2500.00 1.708 1.610
1250.00 1.273 1.175
625.00 0.975 0.877
312.50 0.547 0.449
156.25 0.357 0.259
78.13 0.236 0.138
0.00 0.098 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant sFLT and the recovery rates were calculated by comparing the measured value to the expected amount of sFLT in samples.
Matrix Recovery range Average
serum(n=5) 90-103% 96%
EDTA plasma(n=5) 96-107% 101%
Heparin plasma(n=5) 79-93% 86%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of sFLT and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 79-92% 87-103% 89-98% 80-101%
EDTA plasma(n=5) 83-97% 95-104% 87-98% 95-102%
Heparin plasma(n=5) 82-98% 93-106% 89-97% 85-94%
전체 세부 정보 보기