Cloud-clone
SKU(재고 관리 코드):LMA912Ge
Magnetic Luminex Assay Kit for Acetylcholine (ACH) ,etc.
Magnetic Luminex Assay Kit for Acetylcholine (ACH) ,etc.
Product No.
LMA912Ge
Organism Species
Pan-species (General).
Sample Type
serum, plasma and other biological fluids
Test Method
Competitive Inhibition
Assay Length
1.5h
Detection Range
0.98-1000pg/mL
Sensitivity
The minimum detectable dose of this kit is typically less than 0.327 pg/mL.
UOM
1Plex 2Plex 3Plex 4Plex 5Plex 6Plex 7Plex 8Plex
Specificity
This assay has high sensitivity and excellent specificity for detection of Acetylcholine (ACH) ,etc..
No significant cross-reactivity or interference between Acetylcholine (ACH) ,etc. and analogues was observed.
Recovery
Matrices listed below were spiked with certain level of Acetylcholine (ACH) ,etc. and the recovery rates were calculated by comparing the measured value to the expected amount of Acetylcholine (ACH) ,etc. in samples.
| Matrix | Recovery range (%) | Average(%) |
| serum(n=5) | 91-105 | 98 |
| EDTA plasma(n=5) | 83-98 | 94 |
| heparin plasma(n=5) | 86-95 | 91 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Acetylcholine (ACH) ,etc. were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Acetylcholine (ACH) ,etc. were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Acetylcholine (ACH) ,etc. and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Sample | 1:2 | 1:4 | 1:8 | 1:16 |
| serum(n=5) | 94-102% | 99-105% | 78-99% | 80-95% |
| EDTA plasma(n=5) | 82-94% | 97-105% | 78-97% | 87-94% |
| heparin plasma(n=5) | 93-101% | 78-94% | 93-101% | 78-101% |
Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Reagents and materials provided
| Reagents | Quantity | Reagents | Quantity |
| 96-well plate | 1 | Plate sealer for 96 wells | 4 |
| Pre-Mixed Standard | 2 | Standard Diluent | 1×20mL |
| Pre-Mixed Magnetic beads (22#:ACH) | 1 | Analysis buffer | 1×20mL |
| Pre-Mixed Detection Reagent A | 1×120μL | Assay Diluent A | 1×12mL |
| Detection Reagent B (PE-SA) | 1×120μL | Assay Diluent B | 1×12mL |
| Sheath Fluid | 1×10mL | Wash Buffer (30 × concentrate) | 1×20mL |
| Instruction manual | 1 |
Assay procedure summary
1. Preparation of standards, reagents and samples before the experiment;
2. Add 50μL standard or sample to each well,
add 10μL magnetic beads,and 50μL Detection Reagent A,incubate 60min at 37°C on shaker;
3. Wash plate on magnetic frame for three times;
4. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
5. Wash plate on magnetic frame for three times;
6. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.
| Magazine | Citations |
| Neurosci Bull | Combined administration of D-galactose and aluminium induces Alzheimer-like lesions in brain PubMed: 21614097 |
| Molecular Neurobiology | Neuroprotective effects of simvastatin and Cilostazol in L-methionine-induced vascular dementia in rats pubmed:27544235 |
| Turk Kardiyol Dern Ars. | Effect of statins on sirtuin 1 and endothelial nitric oxide synthase expression in young patients with a history of premature myocardial infarction Pubmed:29664427 |
| JOURNAL OF LIPID RESEARCH | Carotid baroreceptor stimulation in obese rats affects white and brown adipose tissues differently in metabolic protection Pubmed: 31126973 |
| NEUROPHARMACOLOGY | Caffeic acid phenethyl ester counteracts doxorubicin-induced chemobrain in Sprague-Dawley rats: Emphasis on the modulation of oxidative stress and … Pubmed: 33011510 |
| FASEB JOURNAL | Irradiation of carotid baroreceptor with low‐intensity pulsed ultrasound exerts different metabolic protection in perirenal, epididymal white adipose tissue and … Pubmed: 32954574 |
| Exp Eye Res | Autonomic nervous system receptor-mediated regulation of mast cell degranulation modulates the inflammation after corneal epithelial abrasion Pubmed:35421396 |
| SAUDI J BIOL SCI | Aqueous Ajwa dates seeds extract improves memory impairment in type-2 diabetes mellitus rats by reducing blood glucose levels and enhancing brain … Pubmed:35531250 |
| Arq. Bras. Med. Vet. Zootec | Study on the mechanisms of Suaeda rigida polysaccharides on the heart inhibition and skeletal muscle promotion in the frog |
| Catalog No. | Related products for research use of Pan-species (General) Organism species | Applications (RESEARCH USE ONLY!) |
| CPA912Ge11 | BSA Conjugated Acetylcholine (ACH) | Immunogen; SDS-PAGE; WB. |
| CPA912Ge21 | OVA Conjugated Acetylcholine (ACH) | Immunogen; SDS-PAGE; WB. |
| PAA912Ge01 | Polyclonal Antibody to Acetylcholine (ACH) | ELISA, CLIA. / IHC-Fr, ICC, IP (predicted). |
| LAA912Ge71 | Biotin-Linked Polyclonal Antibody to Acetylcholine (ACH) | WB; IHC; ICC. |
| CEA912Ge | ELISA Kit for Acetylcholine (ACH) | Enzyme-linked immunosorbent assay for Antigen Detection. |
| LMA912Ge | Magnetic Luminex Assay Kit for Acetylcholine (ACH) ,etc. | Magnetic Luminex Assay for Antigen Detection. |
| KSA912Ge11 | ELISA Kit DIY Materials for Acetylcholine (ACH) | Main materials for "Do It (ELISA Kit) Yourself". |
Share