Magnetic Luminex Assay Kit for Angiogenin (ANG) ,etc.

Magnetic Luminex Assay Kit for Angiogenin (ANG) ,etc.

Detection range
Size
Sensitivity

Product No.

LMA007Mu

Organism Species

Mus musculus (Mouse).

Sample Type

serum, plasma and other biological fluids

Test Method

Double-antibody Sandwich

Assay Length

3.5h

Detection Range

0.02-20ng/mL

Sensitivity

The minimum detectable dose of this kit is typically less than 0.007 ng/mL.

UOM

1Plex 2Plex 3Plex 4Plex 5Plex 6Plex 7Plex 8Plex

 

Specificity

This assay has high sensitivity and excellent specificity for detection of Angiogenin (ANG) ,etc..
No significant cross-reactivity or interference between Angiogenin (ANG) ,etc. and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Angiogenin (ANG) ,etc. and the recovery rates were calculated by comparing the measured value to the expected amount of Angiogenin (ANG) ,etc. in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-92 81
EDTA plasma(n=5) 88-103 101
heparin plasma(n=5) 78-91 83

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Angiogenin (ANG) ,etc. were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Angiogenin (ANG) ,etc. were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Angiogenin (ANG) ,etc. and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 86-101% 86-93% 80-91% 98-105%
EDTA plasma(n=5) 96-103% 93-102% 79-97% 88-98%
heparin plasma(n=5) 94-105% 79-102% 97-105% 80-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:ANG) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

 

Magazine Citations
Inflammopharmacology Evening primrose oil and celecoxib inhibited pathological angiogenesis, inflammation, and oxidative stress in adjuvant-induced arthritis: novel role of angiopoietin-1 Pubmed: 24664592
Cancer Cell Angiogenin/Ribonuclease 5 is an EGFR ligand and a serum biomarker for erlotinib sensitivity in pancreatic cancer Pubmed:29606349
Journal of Immunotherapy The VE-PTP Inhibitor AKB-9778 Improves Antitumor Activity and Diminishes the Toxicity of Interleukin 2 (IL-2) Administration Pubmed: 31348125
Clinical Hemorheology and Microcirculation Nonselective alpha-/beta-AR antagonists can inhibit pericyte proliferation, migration, and secretion in vitro Pubmed: 32224529
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPA007Mu01 Recombinant Angiogenin (ANG) Positive Control; Immunogen; SDS-PAGE; WB.
RPA007Mu02 Recombinant Angiogenin (ANG) Positive Control; Immunogen; SDS-PAGE; WB.
PAA007Mu01 Polyclonal Antibody to Angiogenin (ANG) WB; IHC; ICC; IP.
PAA007Mu02 Polyclonal Antibody to Angiogenin (ANG) WB; IHC; ICC; IP.
SEA007Mu ELISA Kit for Angiogenin (ANG) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA007Mu Magnetic Luminex Assay Kit for Angiogenin (ANG) ,etc. Magnetic Luminex Assay for Antigen Detection.