제품 정보로 건너뛰기
1 1

Cloud-clone

SKU(재고 관리 코드):LMA460Ge

Magnetic Luminex Assay Kit for Epitestosterone (ET) ,etc.

Magnetic Luminex Assay Kit for Epitestosterone (ET) ,etc.

Detection range
Size
Sensitivity

Product No.

LMA460Ge

Organism Species

Pan-species (General).

Sample Type

Serum, plasma and other biological fluids

Test Method

Competitive Inhibition

Assay Length

1.5h

Detection Range

9.77-10000pg/mL

Sensitivity

The minimum detectable dose of this kit is typically less than 3.257 pg/mL.

UOM

1Plex 2Plex 3Plex 4Plex 5Plex 6Plex 7Plex 8Plex

 

Specificity

This assay has high sensitivity and excellent specificity for detection of Epitestosterone (ET) ,etc..
No significant cross-reactivity or interference between Epitestosterone (ET) ,etc. and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of Epitestosterone (ET) ,etc. and the recovery rates were calculated by comparing the measured value to the expected amount of Epitestosterone (ET) ,etc. in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 96-104 101
EDTA plasma(n=5) 99-105 102
heparin plasma(n=5) 89-104 92

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Epitestosterone (ET) ,etc. were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Epitestosterone (ET) ,etc. were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Epitestosterone (ET) ,etc. and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 84-99% 94-105% 92-99% 78-97%
EDTA plasma(n=5) 85-99% 90-103% 86-94% 79-99%
heparin plasma(n=5) 78-104% 96-103% 86-101% 95-104%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:ET) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 50μL standard or sample to each well,
    add 10μL magnetic beads,and 50μL Detection Reagent A,incubate 60min at 37°C on shaker;
3. Wash plate on magnetic frame for three times;
4. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
5. Wash plate on magnetic frame for three times;
6. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

 

Magazine Citations
Applied Biological Chemistry Positive enhancement of Lactobacillus fermentum HY01 on intestinal movements of mice having constipation 10.1007/s13765-017-0327-3
Nutrients Rotundic Acid Protects against Metabolic Disturbance and Improves Gut Microbiota in Type 2 Diabetes Rats Pubmed: 31887996
OBESITY SURGERY The Protective Effects of Butorphanol on Pulmonary Function of Patients with Obesity Undergoing Laparoscopic Bariatric Surgery: a Double-Blind Randomized … Pubmed: 32535786
mannose-binding lectin reduces oxidized low_x0002_density lipoprotein induced vascular endothelial cells injury by modulating autophagy via LOX1
Evidence-Based Complementary and Alternative Medicine Pretreatment with Shenmai Injection Protects against Coronary Microvascular Dysfunction Pubmed:35722150
Catalog No. Related products for research use of Pan-species (General) Organism species Applications (RESEARCH USE ONLY!)
CEA460Ge ELISA Kit for Epitestosterone (ET) Enzyme-linked immunosorbent assay for Antigen Detection.
IEA460Ge Instant ELISA Kit for Epitestosterone (ET) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA460Ge Magnetic Luminex Assay Kit for Epitestosterone (ET) ,etc. Magnetic Luminex Assay for Antigen Detection.
전체 세부 정보 보기