제품 정보로 건너뛰기
1 1

Cloud-clone

SKU(재고 관리 코드):HES103Hu

High Sensitive ELISA Kit for Complement Component 4d (C4d)

High Sensitive ELISA Kit for Complement Component 4d (C4d)

UOM
Sensitivity
Detection range

Enzyme-linked immunosorbent assay for Antigen Detection.

Product No.

HES103Hu

Organism Species

Homo sapiens (Human).

Sample Type

serum, plasma and other biological fluids

Test Method

Double-antibody Sandwich

Assay Length

3h

Detection Range

9.4-600pg/mL

Sensitivity

The minimum detectable dose of this kit is typically less than 3.2pg/mL.

UOM

48T 96T 96T*5 96T*10 96T*100

 

Specificity

This assay has high sensitivity and excellent specificity for detection of High Sensitive Complement Component 4d (C4d).
No significant cross-reactivity or interference between High Sensitive Complement Component 4d (C4d) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant High Sensitive Complement Component 4d (C4d) and the recovery rates were calculated by comparing the measured value to the expected amount of High Sensitive Complement Component 4d (C4d) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 96-103 99
EDTA plasma(n=5) 93-101 98
heparin plasma(n=5) 78-88 82

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level High Sensitive Complement Component 4d (C4d) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level High Sensitive Complement Component 4d (C4d) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of High Sensitive Complement Component 4d (C4d) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 99-105% 84-92% 78-91% 96-105%
EDTA plasma(n=5) 88-97% 97-105% 85-97% 81-94%
heparin plasma(n=5) 92-99% 97-105% 87-101% 90-104%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

Magazine Citations
Am J Clin Pathol Measurement of Serum Levels of Complement Component 4d as a Quantitative Test of Antibody-Mediated Rejection Content: 144
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPS103Hu01 Recombinant Complement Component 4d (C4d) Positive Control; Immunogen; SDS-PAGE; WB.
PAS103Hu01 Polyclonal Antibody to Complement Component 4d (C4d) WB; IHC; ICC; IP.
MAS103Hu22 Monoclonal Antibody to Complement Component 4d (C4d) WB; IHC; ICC; IP.
MAS103Hu23 Monoclonal Antibody to Complement Component 4d (C4d) WB; IHC; ICC; IP.
MAS103Hu21 Monoclonal Antibody to Complement Component 4d (C4d) WB; IHC; ICC; IP.
SES103Hu ELISA Kit for Complement Component 4d (C4d) Enzyme-linked immunosorbent assay for Antigen Detection.
HES103Hu High Sensitive ELISA Kit for Complement Component 4d (C4d) Enzyme-linked immunosorbent assay for Antigen Detection.
SCS103Hu CLIA Kit for Complement Component 4d (C4d) Chemiluminescent immunoassay for Antigen Detection.
LMS103Hu Magnetic Luminex Assay Kit for Complement Component 4d (C4d) ,etc. Magnetic Luminex Assay for Antigen Detection.
전체 세부 정보 보기