SKU(재고 관리 코드):CEA825Hu
ELISA Kit for Ischemia Modified Albumin (IMA)
ELISA Kit for Ischemia Modified Albumin (IMA)
Enzyme-linked immunosorbent assay for Antigen Detection.
Product No.
CEA825Hu
Organism Species
Homo sapiens (Human).
Sample Type
serum, plasma and other biological fluids
Test Method
Competitive Inhibition
Assay Length
2h
Detection Range
37.0-3,000ng/mL
Sensitivity
The minimum detectable dose of this kit is typically less than 17.3ng/mL.
UOM
48T 96T 96T*5 96T*10 96T*100
Specificity
This assay has high sensitivity and excellent specificity for detection of Ischemia Modified Albumin (IMA).
No significant cross-reactivity or interference between Ischemia Modified Albumin (IMA) and analogues was observed.
Recovery
Matrices listed below were spiked with certain level of recombinant Ischemia Modified Albumin (IMA) and the recovery rates were calculated by comparing the measured value to the expected amount of Ischemia Modified Albumin (IMA) in samples.
Matrix | Recovery range (%) | Average(%) |
serum(n=5) | 90-98 | 94 |
EDTA plasma(n=5) | 81-92 | 84 |
heparin plasma(n=5) | 80-92 | 88 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Ischemia Modified Albumin (IMA) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Ischemia Modified Albumin (IMA) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Ischemia Modified Albumin (IMA) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample | 1:2 | 1:4 | 1:8 | 1:16 |
serum(n=5) | 82-101% | 79-95% | 84-101% | 78-98% |
EDTA plasma(n=5) | 98-105% | 78-97% | 79-93% | 82-91% |
heparin plasma(n=5) | 78-99% | 90-103% | 86-99% | 97-105% |
Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Reagents and materials provided
Reagents | Quantity | Reagents | Quantity |
Pre-coated, ready to use 96-well strip plate | 1 | Plate sealer for 96 wells | 4 |
Standard | 2 | Standard Diluent | 1×20mL |
Detection Reagent A | 1 | Assay Diluent A | 1×12mL |
Detection Reagent B | 1×120µL | Assay Diluent B | 1×12mL |
Reagent Diluent | 1×300µL | Stop Solution | 1×6mL |
TMB Substrate | 1×9mL | Instruction manual | 1 |
Wash Buffer (30 × concentrate) | 1×20mL |
Assay procedure summary
1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
And then add 50µL prepared Detection Reagent A immediately.
Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.
Magazine | Citations |
Aging Clin Exp Res | The role of protein oxidation and DNA damage in elderly hypertension PubMed: 26487663 |
The Anatolian Journal of Cardiology | The impact of protein oxidation on sustained and white coat hypertension pubmed:27684518 |
Indian Journal of Thoracic and Cardiovascular Surgery | Investigation of serum ischemia-modified albumin levels in coronary artery disease patients |
Journal of Cardiovascular Medicine | Effects of Different Heart Positions During Off-pump Coronary Artery Bypass Surgery on Serum Ischemia Modified Albumin Level and Cardiac Output |
Catalog No. | Related products for research use of Homo sapiens (Human) Organism species | Applications (RESEARCH USE ONLY!) |
NPA825Hu01 | Native Ischemia Modified Albumin (IMA) | Positive Control; Immunogen; SDS-PAGE; WB. |
PAA825Hu01 | Polyclonal Antibody to Ischemia Modified Albumin (IMA) | WB; IHC; ICC; IP. |
LAA825Hu91 | HRP-Linked Polyclonal Antibody to Ischemia Modified Albumin (IMA) | WB; IHC; ICC. |
SEA825Hu | ELISA Kit for Ischemia Modified Albumin (IMA) | Enzyme-linked immunosorbent assay for Antigen Detection. |
CEA825Hu | ELISA Kit for Ischemia Modified Albumin (IMA) | Enzyme-linked immunosorbent assay for Antigen Detection. |
CCA825Hu | CLIA Kit for Ischemia Modified Albumin (IMA) | Chemiluminescent immunoassay for Antigen Detection. |
LMA825Hu | Magnetic Luminex Assay Kit for Ischemia Modified Albumin (IMA) ,etc. | Magnetic Luminex Assay for Antigen Detection. |
KSA825Hu11 | ELISA Kit DIY Materials for Ischemia Modified Albumin (IMA) | Main materials for "Do It (ELISA Kit) Yourself". |