1 1

Cloud-clone

SKU(재고 관리 코드):CEX126Ge

ELISA Kit for Folinic Acid (FA)

ELISA Kit for Folinic Acid (FA)

UOM
Sensitivity
Detection range

Enzyme-linked immunosorbent assay for Antigen Detection.

Product No.

CEX126Ge

Organism Species

Pan-species (General).

Sample Type

Serum, plasma and other biological fluids

Test Method

Competitive Inhibition

Assay Length

2h

Detection Range

117.2-30,000pg/mL

Sensitivity

The minimum detectable dose of this kit is typically less than 41.5pg/mL.

UOM

48T 96T 96T*5 96T*10 96T*100

 

Specificity

This assay has high sensitivity and excellent specificity for detection of Folinic Acid (FA).
No significant cross-reactivity or interference between Folinic Acid (FA) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of Folinic Acid (FA) and the recovery rates were calculated by comparing the measured value to the expected amount of Folinic Acid (FA) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 90-104 96
EDTA plasma(n=5) 98-105 102
heparin plasma(n=5) 88-102 95

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Folinic Acid (FA) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Folinic Acid (FA) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Folinic Acid (FA) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 88-98% 99-105% 78-99% 90-102%
EDTA plasma(n=5) 94-101% 79-103% 79-90% 79-94%
heparin plasma(n=5) 83-91% 80-88% 81-95% 98-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.

Catalog No. Related products for research use of Pan-species (General) Organism species Applications (RESEARCH USE ONLY!)
CEX126Ge ELISA Kit for Folinic Acid (FA) Enzyme-linked immunosorbent assay for Antigen Detection.
LMX126Ge Magnetic Luminex Assay Kit for Folinic Acid (FA) ,etc. Magnetic Luminex Assay for Antigen Detection.