SKU(재고 관리 코드):ELK9425
Pig BMG/b2-MG(Beta-2-Microglobulin) ELISA Kit
Pig BMG/b2-MG(Beta-2-Microglobulin) ELISA Kit
Pig BMG/b2-MG(Beta-2-Microglobulin) ELISA Kit
Overview
Product name: | Pig BMG/b2-MG(Beta-2-Microglobulin) ELISA Kit |
Reactivity: | Pig |
Alternative Names: | Beta 2M; BMG; B2MG; b2M; Beta-2-Microglobulin |
Assay Type: | Sandwich |
Sensitivity: | 1.19 ng/mL |
Standard: | 200 ng/mL |
Detection Range: | 3.13-200 ng/mL |
Sample Type: | Serum, plasma, urine and other biological fluids |
Assay Length: | 3.5h |
Research Area: | Infection immunity;Kidney biomarker; |
Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Pig BMG/b2-MG. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Pig BMG/b2-MG. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Pig BMG/b2-MG, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Pig BMG/b2-MG in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
200.00 | 2.204 | 2.109 |
100.00 | 1.602 | 1.507 |
50.00 | 1.276 | 1.181 |
25.00 | 0.826 | 0.731 |
12.50 | 0.567 | 0.472 |
6.25 | 0.335 | 0.240 |
3.13 | 0.226 | 0.131 |
0.00 | 0.095 | 0.000 |
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant BMG/b2-MG and the recovery rates were calculated by comparing the measured value to the expected amount of BMG/b2-MG in samples.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 85-101% | 93% |
EDTA plasma(n=5) | 78-92% | 85% |
Heparin plasma(n=5) | 85-97% | 91% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of BMG/b2-MG and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 87-101% | 89-99% | 85-95% | 93-102% |
EDTA plasma(n=5) | 93-106% | 87-102% | 84-98% | 87-102% |
Heparin plasma(n=5) | 92-102% | 81-93% | 86-98% | 89-97% |