size
Sensitivity
Range

Delve into the intricate world of hormone research with our Rabbit E2 (Estradiol) ELISA Kit. Crafted for sensitivity and precision, this kit enhances your research in reproductive biology, endocrinology, and beyond. Here are the key features of this kit: Size and Sensitivity: Each kit includes materials for 96 tests, ensuring you have a sufficient supply for your research. With high sensitivity, it accurately measures estradiol levels in rabbit samples. Precision and Specificity: Our kit is designed with precision and specificity in mind, ensuring accurate results and minimizing cross-reactivity. Reproductive and Endocrine Research: This kit is your ideal companion for studying estradiol levels, reproductive health, and endocrine system function in rabbits. Quality Assurance: Rigorous quality control measures are implemented to maintain consistent and reliable performance. Explore the dynamics of estradiol in rabbits with the sensitivity and accuracy of our Rabbit E2 (Estradiol) ELISA Kit.

Product Name

Rabbit E2(Estradiol)ELISA Kit

Alias

E2 ELISA Kit, Estradiol ELISA Kit

Application

E2ELISA Kit allows for the in vitro quantitative determination of E2 concentrations in serum, plasma, tissue homogenates and other biological fluids.

Size

96T

Range

12.5-800pg/ml

Sensitivity

< 7.5pg/ml

Species

Rabbit

Standard Curve

Storage

2-8 ℃ for 6 months

Recovery

Matrices listed below were spiked with certain level of E2 and the recovery rates were calculated by comparing the measured value to the expected amount of E2 in samples.

Matrix

Recovery range(%)

Average(%)

serum(n=5)

88-105

97

EDTA plasma(n=5)

90-104

98

heparin plasma(n=5)

85-105

99

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of E2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample

1:2

1:4

1:8

serum(n=5)

85-98%

90-105%

89-100%

EDTA plasma(n=5)

85-94%

90-101%

82-97%

heparin plasma(n=5)

82-99%

80-98%

88-99%

CV(%)

Intra-Assay: CV<8%
Inter-Assay: CV<10%

Note

For Research Use Only

 

전체 세부 정보 보기