제품 정보로 건너뛰기
1 1

FineTest

SKU(재고 관리 코드):ER1507

Rat E2(Estradiol) ELISA Kit

Rat E2(Estradiol) ELISA Kit

size
Sensitivity
Range

Product Name

Rat E2(Estradiol) ELISA Kit

Alias

E2 ELISA Kit, Estradiol ELISA Kit

Detection Method

Competitive ELISA, Coated with Antigen

Application

E2 ELISA Kit allows for the in vitro quantitative determination of E2 concentrations in serum, plasma, tissue homogenates and other biological fluids.

Size

96T

Range

12.5-800pg/ml

Sensitivity

< 7.5pg/ml

Species

Rat

Standard Curve

Storage

2-8 ℃ for 6 months

Recovery

Matrices listed below were spiked with certain level of E2 and the recovery rates were calculated by comparing the measured value to the expected amount of E2 in samples.

Matrix

Recovery range(%)

Average(%)

serum(n=5)

88-103

99

EDTA plasma(n=5)

86-97

92

heparin plasma(n=5)

89-103

97

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of E2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample

1:2

1:4

1:8

serum(n=5)

88-95%

86-92%

85-101%

EDTA plasma(n=5)

82-100%

82-98%

86-98%

heparin plasma(n=5)

81-93%

90-99%

82-93%

CV(%)

Intra-Assay: CV<8%

Inter-Assay: CV<10%

Note

For Research Use Only

 

전체 세부 정보 보기