Overview
| Product name: | Rat sIL-6R(SolubleInterleukin 6 Receptor) ELISA Kit |
| Reactivity: | Rat |
| Alternative Names: | IL-6R;CD126; gp80; IL6R; IL-6R-1; IL-6R-Alpha; IL6RA; IL-6 receptor subunit alpha; Membrane glycoprotein 80; IL6R; Interleukin 6 Receptor |
| Assay Type: | Sandwich |
| Sensitivity: | 0.061 ng/mL |
| Standard: | 10 ng/mL |
| Range: | 0.16-10 ng/mL |
| Sample Type: | serum, plasma, tissue homogenates and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Signal transduction;CD & Adhesion molecule;Infection immunity; |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat sIL-6R. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat sIL-6R. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat sIL-6R, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat sIL-6R in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.166 | 2.062 |
| 5.00 | 1.741 | 1.637 |
| 2.50 | 1.154 | 1.050 |
| 1.25 | 0.886 | 0.782 |
| 0.63 | 0.563 | 0.459 |
| 0.32 | 0.388 | 0.284 |
| 0.16 | 0.175 | 0.071 |
| 0.00 | 0.104 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant sIL-6R and the recovery rates were calculated by comparing the measured value to the expected amount of sIL-6R in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-93% | 86% |
| EDTA plasma(n=5) | 94-105% | 98% |
| Heparin plasma(n=5) | 95-107% | 101% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of sIL-6R and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-105% | 82-101% | 91-103% | 95-102% |
| EDTA plasma(n=5) | 87-91% | 85-93% | 89-97% | 85-93% |
| Heparin plasma(n=5) | 89-97% | 85-94% | 87-96% | 85-97% |