제품 정보로 건너뛰기
1 1

ELK Biotechnology

SKU(재고 관리 코드):ELK9404

Zebrafish GH(Growth Hormone) ELISA Kit

Zebrafish GH(Growth Hormone) ELISA Kit

Sensitivity

Zebrafish GH(Growth Hormone) ELISA Kit

Overview

Product name: Zebrafish GH(Growth Hormone) ELISA Kit
Reactivity: Zebrafish
Alternative Names: GH1; GH-N; GHN; hGH-N; Somatotropin; Hygetropin; Jintropin; Kigtropin; Pituitary Growth Hormone; Growth Hormone, Normal
Assay Type: Competitive Inhibition
Sensitivity: 55.7 pg/mL
Standard: 10000 pg/mL
Detection Range: 156.25-10000 pg/mL
Sample Type: serum, plasma and other biological fluids
Assay Length: 2h
Research Area: Endocrinology;Neuro science;Hormone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Zebrafish GH protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Zebrafish GH. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Zebrafish GH in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
10000.00 0.198
5000.00 0.415
2500.00 0.596
1250.00 0.842
625.00 1.275
312.50 1.537
156.25 1.769
0.00 2.213

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant GH and the recovery rates were calculated by comparing the measured value to the expected amount of GH in samples.
Matrix Recovery range Average
serum(n=5) 82-95% 88%
EDTA plasma(n=5) 78-96% 87%
Heparin plasma(n=5) 87-99% 93%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of GH and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 91-104% 79-94% 87-96% 82-95%
EDTA plasma(n=5) 89-102% 85-92% 96-104% 97-106%
Heparin plasma(n=5) 86-97% 92-99% 85-94% 97-106%
전체 세부 정보 보기